UltiMate 3000 Series:
AFC-3000 Automated Fraction Collector
Page
34
Operating Instructions
Fig. 12: Diverter valve assembly with drop former
4.
Slide the rinse station up or down in its holder so that the lowest part of the drop
former is slightly above the top of the rinse station when it is in the rinse position.
Tip:
This setting ensures that the diverter valve assembly will not impinge on
the rinse station during movement, while allowing for the minimum
distance between those two components to minimize the risk of eluent
being dispensed outside of the rinse station barrel. In case of an
obstruction during the movement of the drop former, the device will turn
off after ca. 20 seconds to avoid damage.
5.
Turn the fraction collector off and manually move the diverter valve assembly to the
four corner positions of the installed fraction racks. This is to verify that the adjusted
height is suitable for all positions despite any tolerances that may exist.
Diverter valve
Drop former
Adjustment
locking lever
Summary of Contents for AFC-3000
Page 2: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Operating Instructions ...
Page 4: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page II Operating Instructions ...
Page 18: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 12 Operating Instructions ...
Page 28: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 22 Operating Instructions ...
Page 48: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 42 Operating Instructions ...
Page 76: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 70 Operating Instructions ...
Page 78: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 72 Operating Instructions ...
Page 82: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 76 Operating Instructions ...