UltiMate 3000 Series:
AFC-3000 Automated Fraction Collector
Page
28
Operating Instructions
3.3.2 Diverter Valve Connections
When connecting capillaries to the fraction collector, observe the following general
precautions:
•
Observe the precautionary statements for capillaries and capillary connections in
section 1.2.2 (
→
page 3).
•
When you connect capillaries, make sure that the connectors are free from contaminants.
Even minute particles may cause damage to the system.
•
Use only the capillaries shipped with the fraction collector and original Dionex spare
capillaries.
Connect the waste and sample inlet tubing to the diverter valve. Verify that the tubings are
properly connected and tightened.
Fig. 5: Connections on the diverter valve
Note the following when connecting the sample inlet:
In order to make sure that the capillary end is not pushed into the valve bore, place the ferrule
at the capillary end so that the capillary does not protrude from the ferrule (
→
Fig. 6). Then
carefully install the capillary.
right
wrong
Fig. 6: Connecting the sample inlet tubing to the diverter valve
Sample inlet
Waste
Summary of Contents for AFC-3000
Page 2: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Operating Instructions ...
Page 4: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page II Operating Instructions ...
Page 18: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 12 Operating Instructions ...
Page 28: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 22 Operating Instructions ...
Page 48: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 42 Operating Instructions ...
Page 76: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 70 Operating Instructions ...
Page 78: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 72 Operating Instructions ...
Page 82: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 76 Operating Instructions ...