UltiMate 3000 Series:
AFC-3000 Automated Fraction Collector
Page
68
Operating Instructions
5.
Loosen the two finger-tight knurled screws that attach the rear cable and grounding
cable to the diverter valve assembly (
→
Fig. 29: Cable screws on the diverter valve assembly
6.
Disconnect the diverter valve cable from the auxiliary I/O connection on the
AFC-3000 rear panel (
→
7.
Loosen the finger-tight knurled screw that attaches the rear cable to the mounting
block on the AFC-3000 rear panel (
→
Fig. 30: Removing the rear cable screw
8.
Reinstall the new diverter valve in reverse order. If you are replacing a low-flow
diverter valve, also see section 7.5 (
→
page 66).
Ground cable
screw
Rear cable screw