Troubleshooting
Detector-Related Problems
Thermo Scientific
Accela RI Detector Hardware Manual
31
Excessive baseline drift Flowcell contaminated
Flush the flowcell with cleaning solvents as described
in
. Check for
leaks.
Mobile phase contamination
Replace the mobile phase with fresh mobile phase
made with high-purity solvents.
Material bleeding from the column Clean or replace the column.
Leaks in system or flowcell
Check all of the fittings for leaks. Call a Thermo
Fisher Scientific field service engineer to replace the
flowcell.
Tiny bubble trapped in the flowcell Increase the flow rate until the bubble is removed.
Connect a backpressure device to the flowcell outlet
(and check backpressure rating to avoid rupturing the
flowcell).
Excessive temperature fluctuations
Remove the system from drafts. Thermostatically
control the column temperature.
No peaks, or peaks
much smaller than
expected
Lamp not on or defective
Make sure the lamp is on. Run the detector
diagnostic tests to check the lamp. Call a Thermo
Fisher Scientific field service engineer to replace the
lamp.
Integrator input voltage not
matching output voltage
Verify that the integrator is connected to the
appropriate analog output connections on the
detector. Check the attenuation setting on the
integrator.
Insufficient sample reaching the
detector
Check the entire system for leaks. Verify the sample
injection volume.
Broad, tailing peaks
Excessive rise time
Lower the rise time selection.
Poor connection at flowcell inlet
Check the end of the inlet tubing for a clean, flat
surface, free of obstructions.
Detector failing to
power up
Tripped circuit breaker at power
outlet
Reset the circuit breaker.
Blown detector fuse
Replace the fuse. See
Incorrect voltage
Verify that the line voltage is within the specified
range as listed in
.
Power cord not connected
Connect the power cord.
Detector not running
upon injection
Detector not receiving Start signal
Check the connection to the Run contacts on the
back panel.
Table 6.
Detector-related problems (Sheet 3 of 4)
Symptom
Cause
Solution