Installation
Making the Back Panel Connections
16
Accela RI Detector Hardware Manual
Thermo Scientific
Connecting the System Interconnect Cable
The timing of the Accela LC modules is synchronized through a system interconnect cable
(
).
Figure 14.
System interconnect cable
Y
To connect the Accela modules to the system interconnect cable
1. To connect the Accela RI Detector, plug one of the three connectors labeled
DETECTOR to the left, 8-pin receptacle located on the back panel of the Accela RI
Detector as shown in
2. To connect an autosampler, plug the unlabeled connector to the left, 8-pin receptacle
located on the back panel of the autosampler.
3. To connect a pump, plug the connector labeled PUMP to the 8-pin receptacle located on
the back panel of the pump.
DETECT
OR
DETECT
O
R
DETECT
O
R
PUMP
PUMP
A/S
M/S
To MS detector
To autosampler
To detector
To detector
To detector
To pump
To pump
Note
If your LC system contains more than one detector, contact your Thermo
Fisher Scientific field service engineer for instructions on how to interconnect the run
signals for the Accela LC modules.