Table 3 Variable set when changed
(continued)
Variable
Type
Value/Description
ErrorRaised
Enum
•
NoError
– There are no errors
•
InternalError
– An internal application
error has occurred
•
FileNotFound
– The
SetLocalProtocolFilePath file cannot be
found
•
JsonParsingError
– The protocol file is
corrupted
•
FileCannotBeOpened
– The protocol file
cannot be opened
•
CmdFailed
– Generic fail, the last
command issued did not succeed
•
ParameterNotSet
– A parameter has not
been set, protocol not loaded
•
ParameterOutOfRange
– A parameter is
out of range, protocol not loaded
•
KitNotLoaded
– The kit is not loaded,
protocol not loaded
•
KitNotValid
– The kit is not valid, protocol
not loaded
•
InstrumentNotReady
– Changing to
the specified.
ProtocolState
cannot be
performed from the current state
•
RecoveryStepOutOfRange
– The
ProtocolStepNumber
is not a valid step
in the protocol
•
ProtocolListDoesNotExist
-
OPCUAProtocolList.yaml
cannot be
found
•
ProtocolListCannotBeOpen
– Cannot
open
OPCUAProtocolList.yaml
•
ProtocolListParsingError
–
OPCUAProtocolList.yaml
format
invalid
•
ProtocolListIdDoesNotExist
– The
LookupTableId
cannot be found in
OPCUAProtocolList.yaml
KitId
String
The Kit id of the loaded kit
Chapter 4
OPC-UA interface
OPC-UA interface
4
CTS
™
Rotea
™
Counterflow Centrifugation System User Guide
101
Summary of Contents for A44769