Component
Fault
Corrective action
Door
Door won’t unlock
Check instrument status and that the door unlock
button is illuminated
Note:
for instructions on how to open the door
manually to retrieve a Single-Use Kit or perform
maintenance, see “Open the door without power”
Door won’t lock
Depress door fully
Check that the KSingle-Use Kit has been correctly
loaded with tubing retained in tube tracks
Check for obstructions
Pinch valves
Valves not fully closing
Perform Single-Use Kit loading and unloading
sequence multiple times, (see “Kit preparation,
loading and removal”
Valves slow to close
Replace Single-Use Kit and check performance
Loud audible noise when
valves close
Replace Single-Use Kit and check performance
Moisture detector
Leaking Kit
Replace Single-Use Kit
Wipe internal surface of centrifuge housing with a
soft cloth or dry paper towel and restart Instrument.
External spill
Wipe internal surface of centrifuge housing with
a soft cloth or dry paper towel and restart the
instrument.
Concentrate output
volume
Incorrect volume or
highly variable output
volume
Check that your priming sequence eliminates air
from fluid lines in the concentrate output path.
Check pump calibration
Check that there isn't any flow restriction in the
output line.
Note:
If delivering into a syringe, friction between
the syringe barrel and the plunger can affect the
output volume.
Note:
Some faults may be rectified by rebooting the instrument. To do this, turn the power off using
either the mains switch or on/off switch on the rear of the instrument, wait 2 minutes and then turn
power back on as described in “Power instrument “On””