Safety instructions
Labels
Safety label is located on the right side of the instrument.
Emergency procedures
This section describes the process for initiating an emergency shutdown and the consequence of a
power failure.
Instrument emergency stop events
The instrument will automatically stop a running protocol if:
• Mains power is lost
• The Moisture Sensor detects fluid leaking in the centrifuge chamber zone
• The Vibration Sensor detects out of balance condition
• One of the Pressor Sensors detects sustained pressure over the threshold
• An internal error of the centrifuge drive, pump drive or valves is detected.
• Safety relay contacts welded or intermittent
• Door closed sensors opening
• Door latched sensor not true.
8
116
CTS
™
Rotea
™
Counterflow Centrifugation System User Guide