The protocol checksum is an hexidecimal number with 32 characters.
Example:
e152abb59e4949980731c5bbf7194ffb
It can be saved and compared as a string.
The
ProtocolState
report should now report
ProtocolSelected
.
Load kit
The instrument verification process needs the instrument door to be closed and a valid kit to be
detected. The kit ID is read by the instrument when the door is closed.
The
DoorState
is reported
Closed
, and the
KitID
string is supported by
KitState
report that checks the
kit type against that required for the protocol. If the kit ID is valid and aligned to the protocol then the
KitState
is reported as valid.
Transfer metadata
Protocols can be defined with metadata or parameters that allow a qualified recipe to respond to
changing input materials and output targets.
Refer to the sequence diagram, see “MES drives Rotea in response to external systems”
After the protocol information is loaded the instrument will set
ParameterName1
through
ParameterName20
to the names of the variables that need to set in the protocol metadata.
The MES can check
ParameterValid1
to
ParameterValid20
to determine the status of metadata
definition for the protocol. This will advise if that parameter is in use, if it has been set, and if the
setting is valid.
To set the value, use
SetParameter1
through
SetParameter20
and supply an
Int32
value.
Initialize protocol on to instrument
Once the protocol has been specified, kit loaded and any metadata defined within acceptable ranges,
set the
SetProtocolState
to
SendProtocolToRotea
.
The protocol definition is sent to the instrument and
ProtocolState
will report
ProtocolReadyToStart
.
Start/Stop/Pause/Resume protocol
• The MES can
Start
,
Stop
and
Pause/Resume
the instrument with
SetProtocolState
variable.
• The
ProtocolState
will report
ProtocolRunning
.
• These actions are available through the buttons on the instrument when enabled.
• The MES can disable
Start
,
Pause
and
Advance
functions with the
SetSkipButtonEnable
and
SetPlayPauseButtonEnable
. When the buttons are enabled they will follow the configuration in the
protocol file, if they are disabled the buttons on the instrument are disabled.
Chapter 4
OPC-UA interface
MES – Commands and reports overview
4
CTS
™
Rotea
™
Counterflow Centrifugation System User Guide
93