Protocol recovery process
Once the conditions causing the fault have been addressed it may be possible to restart a protocol from
a specific protocol step.
Note:
Unless there has been a breach of the closed fluidic system, the cells will have been retained in
the system and are potentially recoverable.
For manufacturing situations with external run history monitoring, qualified protocols specific to the
point in the program where the triggering event for the shut down occurred, can be initiated.
Where this isn’t possible, manual intervention may be required to re-locate the fluid containing cells into
a known bag and conduct a custom protocol to recover the cells.
If power is not available or the instrument is not serviceable,
describes how to manually
open the door to enable removal of the Single-Use Kit for manual processing or transfer to a second
instrument.
Chapter 8
Safety instructions
Emergency procedures
8
118
CTS
™
Rotea
™
Counterflow Centrifugation System User Guide