3200 Series
User Manual
Part No HA028651 Issue 11.0 July-12 39
8.
Temperature (or Process) Input
Parameters in the input list configure the input to match your sensor. These parameters provide the following features:-
Input Type and
linearisation
Thermocouple (TC) and 3-wire resistance thermometer (RTD) temperature detectors
Linear input (-10 to +80mV). 0-10V using external voltage divider. mA assumes a 2.49
Ω
external shunt.
See the table in section 0. for the list of input types available
Display units and
resolution
The change of display units and resolution will all the parameters related to the process
variable
Input filter
First order filter to provide damping of the input signal. This may be necessary to prevent
the effects of excessive process noise on the PV input from causing poor control and
indication. More typically used with linear process inputs.
Fault detection
Sensor break is indicated by an alarm message ‘
Sbr
’. For thermocouple it detects when
the impedance is greater than pre-defined levels; for RTD when the resistance is less than
12
Ω
.
User calibration
Either by simple offset or by slope and gain. See section 8.2. for further details.
Over/Under range
When the input signal exceeds the input span by more than 5% the PV will flash indicating
under or over range. If the value is too high to fit the number of characters on the display
‘HHHH’ or ‘LLLL’ will flash. The same indications apply when the display is not able to show
the PV, for example, when the input is greater than 999.9
o
C with one decimal point.
8.1
Process Input Parameters
INPUT LIST
I N P U T
Name
Scrolling Display
Parameter Description
Value
Default
Access Level
IN . T Y P
INPUT TYPE
Selects input linearisation and
range
See section 8.1.1. for input types available
Conf
L3 R/O
U N IT S
DISPLAY UNITS
Display units shown on the
instrument
none
No units - only for custom linearisation
o
C
L3
o
C
Celsius
o
F
Fahrenheit
o
k
Kelvin
PErc
%
D E C . P
DISPLAY POINTS
Decimal point position
nnnn
No DP
nnnn
Conf
L3 R/O
nnn.n
One DP
nn.nn
Two DP
M V. H I
LINEAR INPUT
HIGH
High limit for mV (mA) inputs
-10.00 to +80.00mV
80.00
Conf
M V. Lo
LINEAR INPUT
LOW
Low limit for mV (mA) inputs
-10.00 to +80.00mV
-10.00
Conf
RNG.H I
RANGE HIGH
LIMIT
Range high limit for
thermocouple RTD and mV
inputs
From the high limit of the selected input type to the
‘Low Range Limit’ parameter minus one display unit.
Conf
L3 R/O
RNG.LO
RANGE LOW
LIMIT
Range low limit for thermocouple
RTD and mV inputs
From the low limit of the selected input type to the
‘High Range Limit’ parameter minus one display unit.
Conf
L3 R/O
PV.OFS
PV OFFSET
A simple offset applied to all
input values.
See section 8.2.
Generally one decimal point more than PV
L3
F IL T . T
FILTER TIME
Input filter time
OFF to 100.0 seconds
1.6
L3
C J. typ
CJC TYPE
Configuration of the CJC type
Auto
Automatic
Auto
Conf and if
T/C
L3 R/O
0
o
C
Fixed at 0
o
C
50
o
C
Fixed at 50
o
C
S B . typ
SENSOR BREAK
TYPE
Defines the action which is
applied to the control output if
the sensor breaks (open circuit).
See also section 8.1.2
oFF
No sensor break will be detected
on
Conf
L3 R/O
on
Open circuit sensor will be detected
Lat
Latching
C JC . i n
CJC
TEMPERATURE
Temperature measured at the
rear terminal block. Used in the
CJC calculation
Read only
Conf
L3 R/O and
if T/C
P v. i n
PV INPUT VALUE
Current measured temperature
Minimum display to maximum display range
Conf
L3 R/O
Summary of Contents for XST-2-0-12-1V1-E28
Page 16: ...12 FURNACE TROUBLE SHOOTING ...
Page 27: ...ɸ ...
Page 28: ...ɸ ...
Page 29: ......
Page 30: ......
Page 31: ...3216 3208 3204 User Manual 3200 Range Process Controller HA028651 11 Aug 2012 ...
Page 170: ...User Manual 3200 Series 138 Part No HA028651 Issue 11 0 July 12 21 Certificate of Conformity ...
Page 171: ......
Page 173: ...Model 2416 Process Controller User Manual Part No HA025041_12 Date August 2014 ...
Page 174: ...This page is intentionally blank ...
Page 259: ...This page is intentionally blank ...