16.3.2
To Calibrate Thermocouple Input
Thermocouples are calibrated, firstly, by following the previous procedure for the mV ranges, then calibrating the CJC.
Connect a mV source as described in section 16.1.3. Set the mV source to ‘
internal compensation’
for the
thermocouple in use and set the output for
0mV
. Then:-
Operation
Do This
Display View
Additional Notes
Select the Calibration List
header
1.
From any display press
as many
times as necessary until the ‘C A L ’
page
header is displayed.
Select the calibration
phase
2.
Press
to select ‘P H A S E ’
none
p h a s e
Scrolling display ‘
C A L IB R A T IO N
p h a s e
’
Select CJC calibration
3.
Press
▲
or
▼
to select ‘
CJC
’
Calibrate CJC
4.
Press
to select ‘GO’
5.
Press
▲
or
▼
to choose ‘
YES
’
The controller automatically calibrates
to the CJC input at 0mV. The display will
show
busy
then
pass
, (if calibration is
successful) or ‘
FAIL
’
if not. Fail may be
due to an incorrect input mV
Summary of Contents for XST-2-0-12-1V1-E28
Page 16: ...12 FURNACE TROUBLE SHOOTING ...
Page 27: ...ɸ ...
Page 28: ...ɸ ...
Page 29: ......
Page 30: ......
Page 31: ...3216 3208 3204 User Manual 3200 Range Process Controller HA028651 11 Aug 2012 ...
Page 170: ...User Manual 3200 Series 138 Part No HA028651 Issue 11 0 July 12 21 Certificate of Conformity ...
Page 171: ......
Page 173: ...Model 2416 Process Controller User Manual Part No HA025041_12 Date August 2014 ...
Page 174: ...This page is intentionally blank ...
Page 259: ...This page is intentionally blank ...