CL-B101D40HZ Manual (d0098)

Front Panel Buttons

Program Button

The 

P

 button is used to move from one program step to the next. 

When pressed at the same time as the   button, it initiates the  

calibration mode

. When pressed at the same time as the   but-

ton, it initiates the 

setpoint setting mode

.

101 Segment

Bargraph

Up Button

When in the operational display, pressing the   button allows you 

to view the setting of the saved Peak and Valley Values

.

When setting a displayed parameter during programming, the  

 button is used to increase the value of the displayed parameter.

Down Button

When in the operational display, pressing the   button allows you 

to change the

 Brightness Level as well as to view the setting of 

the setpoints 

SP1, SP2, SP3 & SP4

.

When setting a displayed parameter during programming, the  

 button is used to decrease the value of the displayed parameter.

Front Panel LED Display

Annunciator LEDs

The annunciator LEDs indicate the alarm status. They are labeled 

from bottom to top: SP1, SP2, SP3, SP4.

Digital LED Displays

The digital LED displays are used to display the meter input  

signal readings. They also display the programming  settings  

during programming.

Setpoint Indication

The position of setpoints on the bargraph display are indicated by 

an ON or OFF segment dependent on the bargraph display being 

above or below the setpoint. 

Setpoints 

indicated 

by an ON 

Segment

Setpoints indicated 

by an OFF Segment

Controls and Indicators

This symbol represents the 

OPERATIONAL DISPLAY.

This is the PROGRAM button.

This is the UP button.

This is the DOWN button.

When a button is shown, press and 

release it to go onto the next step in the 

direction indicated by the arrow. When 

an  alternative  dotted  line  is  shown,  this 

indicates that an alternative logic branch 

will be followed when a particular option 

is present.

When two buttons are shown side by side 

and enclosed by a dotted line, they must 

be pressed at the same time then released 

to go onto the next programming step.

If an X appears through a digit, it means that 

any number displayed in that digit is not rel-

evant to the function being explained.

P

[Span]

[10000]

P

P

When the    and    buttons are shown 

together, the display value can be increased 

by pressing and releasing the    button  

or decreased by pressing and releasing the 

 button.

When the    and    buttons are shown  

with two displays, either display can be 

selected by pressing and releasing the   or 

 buttons.

When two displays are shown together 

with bursts, this indicates that the display is  

toggling (flashing) between the name of the 

function and the value.

Text  or  numbers  shown  between  square 

brackets  in  a  procedure  indicate  the  pro

-

gramming code name of the function or the 

value displayed on the meter display.

When there are more than two display selec-

tions they are shown in brackets below the 

first display and are also selectable by 

pressing and releasing the   or   buttons.

A dotted line enclosing an entire logic dia-

gram indicates that programming branch 

will appear only when a particular option is 

present.

To explain software programming procedures, logic diagrams are

used to visually assist in following the programming steps. The 

following symbols are used throughout the logic diagrams to 

represent the buttons and indicators on the meter:

Programming Conventions

[X•XXX]

[XX•XX]

[XXX•X]

[XXXX•]

[XXXX]

Front Panel Controls and Indicators

LED

Annunciators

for Setpoints

1-4

Single or 

Tri-color 

Bargraaph

Display

UP

Button

DOWN

Button

PROGRAM

Button

7 Segments 

Digital 

Display

Summary of Contents for CL-B101D40HZ

Page 1: ...Mode Selection 6 Bargraph Color Programming Mode 9 Case Dimensions 13 Component Layout 11 Connector Pinouts and Pin Descriptions 10 11 Controls and Indicators 2 Decimal Point and Brightness Selection...

Page 2: ...irection indicated by the arrow When an alternative dotted line is shown this indicates that an alternative logic branch will be followed when a particular option is present When two buttons are shown...

Page 3: ...See Page 5 Goes directly to Zero setting if Analog Output is not installed 4 00 20 00 Range 99 99 9999 999 9 Select one of three Ranges 99 99Hz 999 9Hz or 9999Hz PROGRAMMABLE OFFSET AND SCALE FACTOR...

Page 4: ...6 meter may be rescaled without applying an external signal by changing the Offset and Scale factor Offset is the reading that the meter will display for a zero input The Offset may be set to any valu...

Page 5: ...this point the display skips directly to STEP G and toggles between Cto and oFF the software is detecting that the optional analog output hardware is NOT installed Digital Span Selection For Bargraph...

Page 6: ...h and AnLo analog low can be reversed to provide a 20 to 4mA output The digital span selected can be as small as two counts when using the analog output to function as a Control or Alarm Driver Small...

Page 7: ...volt output selected However the output of cLo must always be less than the value selected for chi If a reversed analog output is desired the values selected to establish the Digital Span can be reve...

Page 8: ...ress the P button Display toggles between PUM and on or oFF STEP F Select Pump PUM on or oFF 1 Using the and buttons select the Pump to be ON or OFF When PUM is selected ON and SP2 is set at a value h...

Page 9: ...argraph Color when the bar is ABOVE SP2 Setpoint 1 Using the and buttons select the desired bargraph color grn oran or red 2 Press the P button Display toggles between CSP3 and the previous color sett...

Page 10: ...mp Form A relays An analog output module is also shown as installed The CL B101D40HZ uses plug in type screw terminal connectors for all input and output connections The power supply connec tions pins...

Page 11: ...A SP4 SP3 SP2 SP1 SP1 SP2 SP3 SP4 Relay Modules with 4 Isolated 5A Form A Relays Output Carrier Board Output Carrier Board Relay Module Relay Module Input Signal Conditioner Analog Output Module Main...

Page 12: ...ditioning Modules COUPLING INPUT SIGNAL FILTER MAG NAMUR SOURCE SINK LOAD AC 20KHz 2KHz 200Hz OFF DC MAG AC LOGIC GND 24V INPUT High Accuracy Digital Measurement NPN Open Collector Proximity Switch No...

Page 13: ...t the meter from external noise Input output leads should be routed as far away as possible from contactors control relays transformers and other noisy components Shielding cables for input output lea...

Page 14: ...y 50 POWER SUPPLY PS1 85 265VAC 95 300VDC N C PS2 18 48VAC 10 72VDC 40 INPUT MODULES Partial List See www texmate com Unless otherwise specified Texmate will ship all modules precalibrated with factor...

Reviews: