41
Note
When the modulation depth is greater than 100%, the
output cannot exceed ±5VPeak (10k
Ω load).
If an external modulation source is selected, modulation
depth is limited to ± 5V from the MOD INPUT terminal on
the rear panel. For example, if the modulation depth is
set to 100%, then the maximum amplitude is +5V, and
the minimum amplitude is -5V.
4-1-7. Selecting the (AM) Modulation Source
The function generator will accept an internal or external source for AM modulation.
The default source is internal.
Panel Operation
1.
Press the MOD key.
MOD
2.
Press F1 (AM).
AM
F1
3.
Press F1 (Source).
Source
F1
4.
Press F1 (INT) or F2 (EXT) to
select the modulation source.
INT
F1
~
EXT
F2
5.
Press Return to go back to the
previous menu.
Return
External Source
Use the MOD INPUT terminal on the
rear panel when using an external
source.
IN
OUT
Trigger
Counter
Trigger
MOD
Note
If an external modulation source is selected, modulation
depth is limited to ± 5V from the MOD INPUT terminal on
the rear panel. For example, if modulation depth is set to
100%, then the maximum amplitude is +5V, and the
minimum amplitude is -5V.