![TEXIO FGX-2220 Instruction Manual Download Page 156](http://html1.mh-extra.com/html/texio/fgx-2220/fgx-2220_instruction-manual_1097519156.webp)
143
8-7-1. SOURce[1|2]:FUNCtion
Set
Query
Description
The FUNCtion command selects and outputs the selected output.
The User parameter outputs an arbitrary waveform previously set
by the SOURce[1|2]:FUNC:USER command.
Note
If the function mode is changed and the current frequency setting
is not supported by the new mode, the frequency setting will be
altered to next highest value.
Vpp and Vrms or dBm amplitude values may have different
maximum values due to differences such as crest factor. For
example, if a 5Vrms square wave is changed to a sinewave, then
the Vrms is automatically adjusted to 3.536.
The modulation, burst and sweep modes can only be used with
some of the basic waveforms. If a mode is not supported, the
conflicting mode will be disabled. See the table below.
Sine
Square Ramp Pulse Noise ARB
AM
✓
✓
✓
✓
✓
FM
✓
✓
✓
PM
✓
✓
✓
FSK
✓
✓
✓
✓
SUM
✓
✓
✓
✓
✓
SWEEP
✓
✓
✓
BURST
✓
✓
✓
Syntax
SOURce[1|2]:FUNCtion {SINusoid|SQUare|RAMP|
PULSe|NOISe| USER}
Example
SOUR1:FUNC SIN
Sets the output as a sine function.
Query Syntax
SOURce[1|2]:FUNCtion?
Return
Parameter
SIN, SQU, RAMP, PULS,
NOIS, USER
Returns the current output type.
Example
SOUR1:FUNC?
SIN
Current output is sine.
Summary of Contents for FGX-2220
Page 1: ...B71 0406 01 INSTRUCTION MANUAL ARBITRARY FUNCTION GENERATOR FGX 2220...
Page 98: ...85...
Page 137: ...124...
Page 227: ...214 9 2 External Dimensions Figure...