Multi Station
Vicat Test Apparatus
21
3.2.3 Additional heat exchanger / art.no.1.0366.13
Figure 14: additional heat exchanger
The Vicat needle device can be equipped with an additional, second heat exchanger to reach a water tem-
perature of +5°C.
Please open the housing in the front.
(Attention! Disconnect electricity!)
Open front plate (only 4, 6 und 8-station devices) and disconnect temperature control wire from control
cabinet.
Then connect Y-Wire and lead it to the outside underneath the device.
Figure 15: open housing
Pull off hose that leads to the basin at the T-switch on the back side.
Figure 16: pull off hose
Unloose this 6
screws
Disconnect D-Sub
plugs here and
connect delivered
Y-wire in between.
Guide the third
plug of the Y-cable
underneath the
heat exchanger to
the outside.
Close the device.
Push on the ring to
remove the hose.
Pull off hose here