Multi Station
Vicat Test Apparatus
17
Operator level no. 3:
To move into the third operator level, first go into the second operator level as described above (Operator
level no. 2). Then scroll up the parameter list to the highest parameter. Then press the UP button and hold
it pressed for at least 10 seconds. The display will show “PA”. Then press the UP and DOWN buttons at the
same time, and hold them both down for at least 4 seconds. This will take you to the parameter list on the
third operat
or level. The parameters there begin with “A1”. You can then use the UP button to move the list
upward, and the DOWN button to scroll downward in this list. When you press the SET button, this will
display the value of a single parameter. If you then press the UP or the DOWN button, then you can increase
or decrease the value of the respective parameter. When you release all the buttons, the new value will be
saved automatically. If you then take no action and press no buttons for more than 60 seconds, the system
will jump back to the first display.
Changing the parameters beginning with the letter “A…” can modify the characteristics of the system. There-
fore, be very careful when modifying these values.
Selection of “A” parameters
Parameter
Description of the function
Setting range
A1
switch mode contact 1
0: heating contact
/
1: cooling contact
A2
switch mode contact 1
0: heating contact /
1: cooling contact
A8
actual value
– display mode
0: integer / 1: 0,5K /
2: 0,1K
A60
sensor selection
11: PT100 two conductors /
12: PT100 three conductors.
A80
Changing the temperature scale
0: Fahrenheit /
1: Celsius
A82
Standby button
0: inactive / 1: active (2/3/4 other function)
3.1.4 Messages and warnings on the display
Message
Description
Action to be taken
OFF
Standby mode. There are no control
functions.
Switch on by pressing the standby-button
F1L
Sensor fault or short circuit.
Check the sensors and the sensor con-
nection terminals.
F1H
Sensor fault or breakage in the sensor.
Check the sensors and the sensor con-
nection terminals.
The keyboard locking function is active.
See Parameter P19 or A19.
The display is
flashing
Temperature alarm (see A31)
The acoustic
alarm sounds
(buzzer)
Temperature alarm (see A31)
Shut off the acoustic alarm (buzzer) by
pressing the DOWN button.
EP
Data loss in parameter memory (control-
contacts no. 1 and 2 are without current)
Switch the mains power OFF, then ON
again. If this does not solve the problem,
the controller must be repaired.
Sensor error messages are saved and will continue to be shown, even if the cause of the problem has been
corrected. If you acknowledge these error messages by pushing the DOWN button, this will delete the error
message.