Laboratory mortar mixer
1.0206.07
50
Test and measurement report
DIN VDE (Association for Electrical, Electronic & Information Technologies) 57 100 – VDE
0701 DGUV (German Social Accident Insurance) Regulation 3
Acceptance of the test
Testing after repair
Device/type
Article number
Manufacturing number
Mortar mixer
1.0206.07
14 - 2016
Nominal data
Voltage
Frequency
Current
kW
min
-1
Protection
class
400 V
50 Hz
1.5 A
0.4
62/125
I ( VDE 0100 )
Item Test/measurement
TARGET
STATE
CURRENT
STATE
1
Visually inspect protective conductor
O.K.
2
Mains voltage during measurement
V
3
Contact resistance
< 300 m
Ω
m
Ω
4
Insulation resistance
>
= 1.0 M
Ω
M
Ω
5
Alternative leakage current
<
= 3.5 mA
mA
6
Functional testing
O.K.
Measuring instruments
used:
Digital multimeter (Voltcraft)
Device tester according to
0701
Repair findings/replacement parts/comments
Notes
A Use of a RESIDUAL CURRENT CIRCUIT BREAKER with 30 mA rated fault current resolution
is recommended.
B The accident prevention regulation VBG 4 (German Institution for Statutory Accident Insurance and Prevention in
the Administrative Sector) requires that measurements be carried out
according to VDE 0701 (Association for Electrical, Electronic & Information Technologies) items 1-4 with a test
cycle of about 6 months for portable electrical devices.
As manufacturer, we are obliged to inform you of this regulation.
Tests and measurements have been carried
out
Date
Signature