Laboratory mortar mixer
1.0206.07
32
General Setup
screen
The different mixer parameters can be set via this window.
For this you can adjust the operation to a certain extent to protect the drum and paddle.
Water Setup
screen
The target values for the water metering system can be adjusted and tested via this
window.
Some cement types demonstrate extreme behaviour regarding the relationship wa-
ter/consistency. It may happen that the mixing time expires before the desired torque
target value is achieved. In this case, the value under
Initial Water
must be increased.
If the required quantity of water lies at or even below 200 ml, the value in the
Initial Wa-
ter
must be reduced.
By changing the
Pump Curve Factor
(pump capacity curve), additional possibilities for
adapting the water metering for extreme types of cement arise.
Adjust the total mixing time
in CBWD mode
(Default value
180 s
)
To adjust the
buzzer volume
To return to the
Main Setup
screen
To adjust the target
torque in CBWD
mode
(Default value
3.0
)
To tare the
torque sensor
(‘Tare’)
For the fine adjust-
ment of the calibra-
tion value 4
(Default value
1.0
)