Laboratory mortar mixer
1.0206.07
6.
Preparations for start
6.1 Installation site/environmental conditions
Warning
The following limit values apply in this regard:
Ambient temperature
DIN EN 60204-1, 4.4.2
Air humidity
DIN EN 60204-1, 4.4.3
Note
Install the
6.2 Electrical connection
A
CEE 16A
power supply connection box
for the mortar mixer’s power supply connection.
Plug the mortar mixer’s 5-wire cable with the CEE
nection box.
16
Preparations for start-up
Installation site/environmental conditions
The operation of the mortar mixer is
permitted only in dry rooms!
The following limit values apply in this regard:
1, 4.4.2
+5
°
C to
+40
°
C
1, 4.4.3
30% to
95%
Install the mortar mixer on a load-bearing, level base.
Electrical connection
power supply connection box
400 Volt 3Ph+1N+1PE, 50 Hz
for the mortar mixer’s power supply connection.
wire cable with the CEE plug (16A) in this power supply co
The operation of the mortar mixer is
bearing, level base.
400 Volt 3Ph+1N+1PE, 50 Hz
must be used
plug (16A) in this power supply con-