The output ±2 V / ±10 V is connected to the analogue display for the ranges grater or equal
to ±500 µm / ±.020 in. That is ±2 V / ±10 V on full scale.
For the ranges lower or equal to ±200 µm / ±.010 in, the output is unique and gives
±2 V / ±10 V for a maximum value of ±200 µm / ±.010 in.
Range µm
Tension full scale
Sensitivity mV / µm
± 5000
± 2 V
± 10 V
0.4
2
± 2000
± 2 V
± 10 V
1
5
± 500
± 2 V
± 10 V
4
20
± 200
± 2 V
± 10 V
10
50
±
50
± 0.5 V
± 2.5 V
10
50
±
20
± 0.2 V
± 1 V
10
50
±
5
± 0.05 V
± 0.25 V
10
50
The maximum capacitive charge is 40 nF, max. 3 mA.
10 ACCESSIBLE FUNCTIONS BY RS232
Features
Transmission speed:
4800 Baud
Parity:
even
Start bit:
1
Character output:
7 ASCII code bits
Stop bit:
2
Send required information followed by <CR><LF>
Mono-directional mode
By using the standard 47.61046 cable, the change in DTR line status returns the displayed
value.
Bi-directional mode
By using the 47.61049 cable, the instrument can be
controlled from a computer. The following instructions are
at your disposal.
50 ms < t < 500 ms
Data Request (DTR)
Data (RXD)
+
–
OV
+
–
OV
5
Data Request (DTR)
V-(TXD)
Data Out (RXD)
1
9
V+(RTS)
6
Printer SPC or
PC side Sub-D 9 pole male
5
V+ (DTR)
Data In (TXD)
Data Out (RXD)
1
9
V– (RTS)
6
Printer SPC or
PC side Sub-D 9 pole male