Guidance and manufacturer’s declaration –
electromagnetic emissions
The BPM105/BPM205 is intended for use in the electromagnetic environment specified
below. The customer or the user of the BPM105/BPM205 should assure that it is used in such
an environment.
Emissions test
Compliance
Electromagnetic environment
–
guidance
RF emissions
CISPR 11
Group 2
The BPM105/BPM205 must emit
electromagnetic energy in order to perform its
intended function. Nearby electronic equipment
may be affected.
RF emissions
CISPR 11
Class B
The BPM105/BPM205 is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
For Details of essential performance: please refer to section 1.1 Intended Use
Guidance and manufacturer’s declaration –
electromagnetic immunity
The BPM105/BPM205 is intended for use in the electromagnetic environment specified
below. The customer or the user of the BPM105/BPM205 should assure that it is used in such
an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment
–
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, oncrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
Not
applicable
Not
applicable
-
Surge
IEC 61000-4-5
Not
applicable
Not
applicable
-
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Not
applicable
Not
applicable
-
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
44