PROBLEM
SOURCE
SOLUTION
Connection
error (signal
lost)
Change
Indicator
1. Make sure the transmitter is properly
attached to the sensor strip and that the
battery is properly inserted. (see p. 11 of
this guide).
2. Press both buttons on the assembled Change
Indicator simultaneously for two seconds.
a. If there is no light, change the battery and
re-attach the transmitter to the sensor strip.
Press both buttons simultaneously. If there
is a steady light for a few seconds,
everything is OK.
b. If there is a blinking light, check that the
transmitter is properly attached to the
sensor strip. Adjust if needed, then press
both buttons simultaneously. If the light
continues blinking, change the battery.
c. If there is a steady light for a few seconds,
everything is OK.
3. If this does not help, replace the sensor
strip. Press both buttons simultaneously.
If there is a steady light for a few seconds,
everything is OK.
4. If this does not help (no light or blinking
light after correctly attaching to the sensor
strip), please contact TENA SmartCare
Support.