CC-10
INSTRUMENT MANUAL
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:[email protected] web site:www.televac.com
CC-10 MANUAL REV D
Page 26 of 40
Y
TELEVA
Command
Mode
Explanation
Response
Data
Explanation
4
Read SP3L, SP3H
ppse must be no larger than
PPSE, otherwise or the invalid
value causes data error ”0003”
Note: The setpoint value must
be between
1.0E-9 and 9.9E3(Torr or mbar),
or 1.0E-7 and 9.9E5(Pa)
5
Read Analog output set
parameter
1ABB
For Log output mode
A: voltage span for 1 decade
0: 0.5V/decade
1: 1.0V/decade BB: full
scale voltage when A=0
BB is 07 to 10
when A=1, BB is 00 to 03
for Torr and mbar unit, or is
02 to 05 for Pa unit
2000
combined output mode
3000
special mode for remote
display unit