SP1 Set Mode
In this mode value of SP1L and SP1H can be set.
SP1
SP2
SP3
HV
SP2
SP1
SP3
HV
HV
SP2
SP1
SP3
SP3
SP2
SP1
HV
X10
Torr
Torr
Torr
Torr
X10
X10
X10
FUNC
FUNC
FUNC
FUNC
ENT
ENT
ENT
ENT
Select SP1H
Select SP1H
Select SP1L
Select SP1L
SP1 mode : When
“parameter lock” is enabled any parameter change can not be done
and only displaying the current parameters is possible. Make sure if lock is disabled
before the programming. (Refer to 4.2)
SP1 lamp blinks when SP1 setting mode is selected. Characters of “SP 1L” appear in
the main display then the pre set value appears. The first digit of the mantissa blinks.
To change the set value, follow the procedures below.
(1) Select the digit to be changed by horizontal arrow key
[→].
(2) Change the value by vertical arrow key [
↑].
(3) By pushing [ENT] key the unit moves to SP1H set mode.
* If, instead, [FUNC] key is pushed the unit skips SP1H then moves to SP2L
mode. Note that the new set value of SP1L is not stored in this operation.
When SP1L is determined the unit moves to SP1H set mode. The character of
“SP 1H” is shown in the display then the pre-set value is appears. The first digit
will blinks prompting the entry of the new value.
To change the set value, follow the procedures below.
(1) Select the digit to be changed by horizontal arrow key [
→].
(2) Change the value by vertical a
rrow key ↑
(3) By pushing [ENT] key the values of both SP1L and SP1H are registered
and the unit returns to measure mode.
* If, instead, [FUNC] key is pushed the values are not registered and the unit moves
to SP2L mode.