Figure 2-19. 640x Inlet and Sample Line Assembly
12.
Connect the ambient temperature sensor cable to the rear panel of the
instrument and route the other end out of the drilled hole (wall or roof), and
extend it out near the instrument’s inlet.
13.
Install the radiation shield either onto the instrument sample tube or nearby
guard rail using the supplied ratcheting ribbon.
14.
Loosen the retention screw and insert the ambient temperature sensor probe
into the bottom of the radiation shield. Position the end of the probe into the
upper center portion of the radiation shield and tighten down on the retention
screw so that it maintains the proper position within the radiation shield.
Figure 2-20. T640 Rooftop Sample Inlet and Extension Tubing Installation