Teledyne API T640 PM Mass Monitor
77
8. When finished, reassemble optics chamber components;
•
Reinsert the nozzle to the top of the optical cell and tighten the screws
to “hand-tight”. Do Not Overtighten the screws.
•
Install the cup to the bottom of optics chamber;
•
Reconnect tubing from cup in bottom of chamber to RH/T sensor
•
Reinstall the ASC support and the tubing from RH/T sensor to the DFU
filter.
o
When re-installing the ASC support, the tubing from the optical
cell cup to the DFU needs to be routed through the side
(above the diagonal support) of the ASC support.
o
Pay special attention not to place the support on top of any
wires near the rear-panel and CPU board. The harness
should be out of the way, but it is a good idea to check.
•
Reconnect the flow sensor plugs to the front control board as well as
any additional tubing that was unplugged (i.e. the pulsation dampener
to the pump).