Teledyne API T640 PM Mass Monitor
84
APPENDIX
A
–
W
ARRANTY
R
EPAIR
Q
UESTIONNAIRE
(Reference 08343B DCN7303)
CUSTOMER: _______________________________ PHONE: ______________________________
CONTACT NAME: ___________________________ FAX NO.
SITE ADDRESS:
MODEL TYPE: _______________ SERIAL NO.: _________________ FIRMWARE REVISION:
Are there any failure messages?
PLEASE COMPLETE THE FOLLOWING TABLE FOR THE T640:
PARAMETER
RECORDED VALUE
ACCEPTABLE VALUE
Sample Flow
LPM 5.0LPM ±5%
Sample Flow Slope
0.85 – 1.15
Ambient Pressure
/- 10mmHg of audit device
Sample RH
%
≤
50% (
typically ≤35%
controlled by
ASC)
Box Temp
°C 0-60°C
Ambient Temp
°C -40 - 60°C
LED Temp
°C
≤70°C
Sample Temp
°C 0 – 60°C
P3 Value
45 - 55
PMT HV Setting
V 1100 – 2200V
Pump PWM
<80%
640x Valve PWM
<85%
640x Option Bypass Flow
LPM 11.67 (±5%)
640x Option Bypass Flow Slope
0.85 – 1.15
* Used for diagnostics only and should not be set/reset without explicit direction from TAPI Tech Support. If
P3 (Calculated) does not appear in the Dashboard, configure the Dashboard (Section 3.2.9) to display it.
What is the measured Sample flow rate? liters/min
What are the failure symptoms? ________________________________________________________
__________________________________________________________________________________
__________________________________________________________________________________
__________________________________________________________________________________