84
Auto Trigger
Auto Trigger parameters that need to be set up include delay, samples/trigger
and VMC out.
Operating Steps:
1. Press
, then select
【
Trg Src
】
→
【
Auto
】
or press
on the front
panel directly to enable Auto Trigger.
2. Set the delay.
Delay is the waiting time after the trigger signal is sent out and before the
acquiring starts. Press
【
Delay
】
to select Auto or Manual mode. When
choosing Manual mode
, the
Left and Right keys are used to select the
numerical digit to adjust. The Up and Down keys are used to change the
selected digit value.
3. Set the number of samples per trigger.
Press
【
Samples/Trigger
】
to set the sample count. Left and Right keys
are used to switch the number of a numerical value, Up and Down keys are
used to change the selected value.
Sample Count
•
Sample Count indicates the number of samples taken when the
Multimeter receives a Single Trigger event (samples per trigger).
•
The range of sampling points should be between 1 and 599999999.
•
The default value of the Sample Count is 1.
4. Set the VMC Out.
The Multimeter outputs a pulse signal through the VM COMP interface on the
rear panel after the sampling signal is finished. Press
【
VMC Out
】
to choose
Positive or Negative polarity.
Summary of Contents for T3DMM6-5
Page 1: ...User Manual T3DMM6 5 Digital Multimeter T3DMM6 5 SC Digital Multimeter ...
Page 18: ...17 Appearance and Size Diagram 1 4 Appearance and Size ...
Page 26: ...25 User Interface Diagram 1 9 User Interface Measurement Measurement ...
Page 29: ...28 Temperature Measurement For RTD and thermcouple sensors ...