Function Menu
Description
Channel Config
Open/close the channel and set the measurement function, measurement
parameters of specified channel.
Low
Set the low value of scanned channel range.
High
Set the high value of scanned channel range.
Press the [Channel Setup] to enter the channel configuration interface and set the channel switch, function, range and speed..
The range setting is applicable for the following functions: DC/AC Voltage (DCV/ACV), 2/4 Wire Resistance (2W/4W), Capacitance (CAP), Frequency (FRQ).
Table 3. Available Range for different measurement function
Measurement Function
Available Range
DCV/ACV/ FRQ
Auto, 200mV, 2V, 20V, 200V
DCI/ACI
2A (fixed)
2W/4W
Auto, 200Ω, 2kΩ, 20kΩ, 200kΩ, 2MΩ(1MΩ for T3DMM6-5), 10MΩ, 100MΩ
CAP
Auto, 2nF, 20nF, 200nF, 2μF, 20μF, 200μF,
10000μF (2mF, 20mF, 100mF for T3DMM6-5)
The scanner function provides two measurement speeds: Fast (50 reading/s = 20ms / reading) and Slow (5 reading/s = 200ms / reading).
(Fast: 1PLC, Slow: 10 PLC for T3DMM6-5-SC) The speed setting is applicable for the following functions: DC/AC Voltage (DCV/ACV), 2/4
Wire Resistance (2W/4W)
Operating instructions:
• Move the cursor to choose the wanted parameter by direction keys and the background color of cursor’s position turns to gray.
• Select the current item by pressing “OK” key and the background color of the selected item turns to green.
• Set the parameter by up and down direction keys.
• Press “OK” key again to store the setting of the selected item of which the background turns back to gray. Move the cursor and repeat the
prior steps to set the next parameter.
• Press [Done] to save the current settings and return to the higher level menu.
5.Channel range setting
Select [High] or [Low] and then input numerical value by direction keys.
Note: The upper limit value should be always bigger than the lower limit
value.
Scanner Function Menu Description
Summary of Contents for T3DMM6-5
Page 1: ...User Manual T3DMM6 5 Digital Multimeter T3DMM6 5 SC Digital Multimeter ...
Page 18: ...17 Appearance and Size Diagram 1 4 Appearance and Size ...
Page 26: ...25 User Interface Diagram 1 9 User Interface Measurement Measurement ...
Page 29: ...28 Temperature Measurement For RTD and thermcouple sensors ...