107
Chapter 5: General Troubleshooting
The following is a list of issues that may arise while using the Multimeter and
its solutions. Please use the corresponding steps when troubleshooting. If you
are still unable to resolve, please contact Teledyne Test Tools.
➢
If the screen of the Multimeter is still dark with nothing displayed
after pressing the power key:
1. Check if the power has been connected correctly.
2. Check if the main power switch on the back panel has been turned
on.
3. Check if the power fuse has blown. Replace it as desired if it has
blown.
4. Restart the Multimeter after all the above steps are complete.
5. If the instrument still cannot start up properly, please contact
Teledyne Test Tools.
➢
The reading does not change when connecting an AC current
signal:
1. Check if the test lead has been connected to the current jack and LO
jack correctly.
2. Check if the fuse in the current location on the back panel has
blown.
3. Check if the correct measurement connectors have been
selected (DCI or ACI).
4. Check if the input is ACI but the measuring location is DCI.