52
To Measure Resistance
Range:
200Ω, 2kΩ, 20kΩ, 200kΩ, 1MΩ, 10MΩ, 100MΩ
Max Resolution:
100μΩ (in the 200Ω range)
Input Protection:
1000 V protection is available in all ranges and a 10%
Over-range for all ranges. If the reading exceeds the range, “
overload
” will
be displayed.
T3DMM6-5 provides 2-wire and 4-wire resistance measurements. When the
measured resistance is lower than 100kΩ, the 4-wire resistance measurement
is recommended to reduce the measurement error caused by the test lead
resistance and contact resistance between the probe and the testing point.
These two resistances should not be ignored when compared to the measured
resistance.
Operating Steps:
1. Enable 2-wire/4-wire resistance measurement
Press
on the front panel to enter the 2-wire resistance measurement
interface, as shown in Diagram 2- 13.
Diagram 2- 13 2 Wire Resistance Measurement Interface