51
4. Set the filter
Press
【
Filter
】
and choose the filter for the measurement. The instrument
provides three different AC filters, “>3Hz”, “>20Hz” and “>200Hz”. You
should generally select the highest frequency filter whose frequency is less
than that of the signal you are measuring.
5. Set the relative value (Optional operation)
Press
【
Rel
】
to start or stop the Relative math function. When it is
started, the reading displayed is a value which comes from the result of
actual measurement value minus the relative value that has been set. The
default relative value is the measurement value when the function is
started. (For details, please refer to “
Math Functions
” in Chapter 2.)
6. Read the measurement value
The multimeter measures the input signal according to the current
measurement settings and displays the measurement result on the screen.
7. Perform the math operation (advanced)
You can perform the math operations (Statistics, Limit and REL) on every
ACI measurement reading. For details, please refer to “
Math Operations
”
in Chapter 2.
8. Display the graph (advanced)
You can analyse the measurement data by using the “Bar Meter”, “Trend
Char” or “Histogram display”. For details, please refer to “
Display Mode
”.