Operator’s Manual Simple AFG Application
46
Note:
the values refer to the rise and fall time between the 10% and the 90% of the pulse amplitude.
The 0% to 100% transitions will be longer than the set values anyway the graph represents the entered
value as the 0% to 100% transition time.
Note:
Using the Pulse Waveform the following constraints must be met:
-
𝑅𝑖𝑠𝑖𝑛𝑔 𝐸𝑑𝑔𝑒 + 𝐹𝑎𝑙𝑙𝑖𝑛𝑔 𝐸𝑑𝑔𝑒 < 𝑃𝑒𝑟𝑖𝑜𝑑
-
𝑊𝑖𝑑𝑡ℎ < 𝑃𝑒𝑟𝑖𝑜𝑑
Auxiliary Channels