Marker Out
The Marker Out generates a digital pulse synchronous with the waveform or with the modulating
function depending on the Run Mode. To set the Marker Out parameters refer to the Channel Settings.
Marker Out Specification Value
Connector
1 BNC per pair of channels on the Front Panel
Output impedance
50 Ω
Output level (into 50 Ω)
1 V to 2.5 V
Marker out (blue, top) synchronous with the analog signal (red, bottom)
Trigger In
The TRIGGER IN (Trigger In connector on the front panel) allows to control the signal generation when a
channel is in Burst Run Mode or in Sweep Run Mode. Refer to the chapters about the Device Settings to
know how to define the trigger parameters or the Run Mode. In Continuous and Modulation Run Modes
the Trigger In
doesn’t have any effects.
Trigger In Specification Value
Connector
1 BNC on the Front Panel
Input impedance
1 kΩ or 50Ω selectable
Slope/Polarity
Positive or negative selectable