Operator’s Manual Simple AFG Application
68
Channels and Device Setting
Channels Settings
Touching the Setting Tab
or slide left right until you reach the setting page.
The parameters present in Channel settings are descripted in the table below:
Channel Setting
Description
Initial Delay
Set the Initial Delay of the selected channel. The delay range is 0s to 14 s.
Load Impedance
[Ohm]
The instrument applies the appropriate scaling to the output waveform to get the
right amplitude on the defined Load expressed in Ohm.
The impedance range is 1 Ohm to 1 MOhm.
Baseline
Offset[V]
Defines the DC offset value added to the output signal respect to the ground level.
The range is between -6 V to 6 V (-HV option) on 50 Ohm load and it depends on
the selected load impedance (i.e. -12V to 12V into open load).
Polarity
This control allows you to invert the polarity of the output signal relative to its
offset.
High Limit [V]
Sets the maximum voltage that the channel generates. This limit is verified during
the generation, but it doesn’t take
into account the Baseline Offset.
The constraint High Limit > Low Limit must be met. This feature can be used to
ensure the load is not damaged.