Multi-Instrument System
In a Multi-Instrument configuration, a Master device can control every triggering and timing setting in
order to synchronize its operation with that of other Slave devices.
You can connect up to four T3AWG-3358/3258 units to have a system with 32 synchronized analog
channels.
In order to set up a multi-Instrument system you must first:
•
Turn off the instruments;
•
Select the instrument that you want to use as Master. The other units will be considered as
Slaves.
•
Using the Sync cable, connect the Master
Sync Out
connector to the Slave
Sync In
connector
located on the rear panel of the instruments. Then connect the Sync Out of the slave to the Sync
In of the next slave device and so on up to the last slave device.
Sync OUT and Sync IN
connectors
Master instrument