GT SERIES
1 INSTRUMENT, 7 APPLICATIONS
2
67112
Revision 10
1.2.1 Additional Safety Requirement - CSA Only
1.2.2 Certifications and Approvals
The monitor has the following approvals:
WARNING: TO REDUCE THE RISK OF IGNITION OF A FLAMMABLE OR
EXPLOSIVE ATMOSPHERE, BATTERIES MUST BE CHANGED ONLY IN A
LOCATION KNOWN TO BE NON-HAZARDOUS.
WARNING: TO REDUCE RISK OF EXPLOSION, DO NOT MIX OLD
BATTERIES WITH USED BATTERIES OR MIX BATTERIES FROM DIFFERENT
MANUFACTURERS.
WARNING: NEVER ATTEMPR TO RECHARGE NON-RECHARGEABLE
CELLS.
Caution: Not for use in oxygen enriched atmospheres.
Caution: Before each day's usage, sensitivity must be tested on a
known concentration of Methane equivalent to 25% - 50% of full-scale
concentration. Accuracy must be within 0 to +20% of actual. Accuracy
may be corrected by using the GT Calibration Software and manually
applying the gas.
Mark
ATEX
SIRA 05ATEX II 2 G EEx ia d IIB T3
IECEx
SIR05.0006X Ex ia d IIB T3
Ex ia T3 C22.2 No.152
UL 913 Class I, Div 1 Groups C and D.
This equipment is designed and manufactured to protect against other hazards as defined in
paragraph 1.2.7 of Annex II of the ATEX Directive 2014/34/EU.