GT SERIES
1 INSTRUMENT, 7 APPLICATIONS
52
67112
Revision 10
Figure 77: Applied Gas Peak Values
• Left column - gas ranges available in the monitor
• Middle column - concentration of gas being applied
• Right column -peak gas reading
4. Apply the gas using a demand flow regulator.
5. Once the peak reading has stabilized, disconnect the gas.
6. Either exit the Bump Mode and store the values automatically or go back to the CalGas menu
to define and apply gas from another cylinder.
• Exit the Bump Mode - press and hold the EXIT button.
• Return to CalGas selection display - press and hold OK button.
7.2.2 Viewing Bump Test Results
From CalGas set-up screen (shown in
Figure 75: Example CalGas Cylinder
), press and hold LAST
button to view the results of the last 32 bump tests.
The screen shown in
Figure 78: Previously Stored Bump Tests
will be displayed.
Figure 78: Previously Stored Bump Tests
• Use the UP
or DOWN
buttons to highlight the required results.
• Press and hold the NEXT button to select.
Once selected, the stored bump test reading will be shown (as illustrated in
).
Figure 79: Stored Bump Test Readings