Operational Theory
BDS 3000
Teledyne Analytical Instruments
12
Figure 2.4 Typical Purge-down Curve After Air Saturation
S h or t e n i ng t h e ai r exp o s u r e wi l l al l o w a f a s t e r se ns o r r ec o v e r y.
A ty p i c a l BD S se n s o r wi l l re c o v e r t o 1 pp m i n ap p r ox i m a t el y 25
m i nu t e s , t o 10 0 pp b af t e r 80 m i n , an d 10 pp b i n ab ou t 8 ho u r s , af t e r
s u f f e r i n g a t e n m i n u t e ex p os u r e to ai r .
2.2.6 Background gas compatibility
T he BDS oxygen sensor wi l l work in i ner t gas backgr ounds,
i ncludi ng ni t rogen, hydr ogen, ar gon, heli um and ethane. The sensor
out put, however, i s dif fer ent in di ff erent backgr ound gases. For exam pl e,
t he sensor output i n a hydr ogen background is twi ce as much as it woul d
be in a ni tr ogen background. Therefor e, i t is recom mended to cali br ate t he
analyzer wit h an oxygen standar d that has a sim il ar background as the
sam pl e gas. If an oxygen st andar d is unavai l able for a par ti cul ar
background, a Gas F act or which is det er mi ned at TAI coul d be used t o
cor rect the sensor out put in dif f er ent backgr ound (see sect ion 4.3.9) .
Note: the gas flow meter in the analyzer is calibrated for air. The
error for measuring nitrogen is usually negligible. But for
hydrogen, it reads 100% lower. For example, when the
float ball in the flow meter is at 0.5 SLPM, the actual flow
rate of hydrogen is about 1 SLPM.
Summary of Contents for BDS 3000
Page 14: ......
Page 32: ...Operational Theory BDS 3000 Teledyne Analytical Instruments 18...
Page 46: ...Installation BDS 3000 Teledyne Analytical Instruments 32...
Page 76: ...Operation BDS 3000 Teledyne Analytical Instruments 62 Figure 4 2 Analyzer Power up Sequence...
Page 77: ...Oxygen Analyzer Operation Teledyne Analytical Instruments 63 Figure 4 3 Analyzer Span Sequence...
Page 78: ...Operation BDS 3000 Teledyne Analytical Instruments 64 Figure 4 4 Analyzer Zero Sequence...
Page 86: ...Maintenance BDS 3000 Teledyne Analytical Instruments 72...
Page 92: ...Appendix BDS 3000 Teledyne Analytical Instruments 78...
Page 96: ...Appendix MSDS Teledyne Analytical Instruments 82...