Oxygen Analyzer
Operation
Teledyne Analytical Instruments
47
lOffset[2] = 2327
(ADC count of offset of the third
gain of sensor amplifier)
lOffset[3] = 2330
(ADC count of offset of the fourth
gain of sensor amplifier)
Fourth Screen:
tempOffset = 2190
(ADC count of offset of
temperature amplifier)
Current_gain= 1
(current gain of sensor amplifier on
span gas)
4.3.11 Temperature
The temperature of the sensor is displayed on the third screen or
system menu, and is shown in degrees centigrade. If sensor cable is
disconnected, the display will show either “> 50” or “ < 0”.
4.3.12 Extended Analog Output Function
To access this function, press the
SYSTEM
key to enter it’s menu.
Select the
More
function to move on to the second screen of the
system
menu.
Version
Diag Xout
Neg-Y Filter-10 More
Select
Xout
and press
ENTER
to get to the extended output setup
menu.
Output Offst: 10% FS
Neg. Analog Offset
The number on the first line is adjustable using the Up and Down
keys between 0 and 50. This function makes it possible to track negative
readings in the analog output, since the analog outputs do not go below
zero volts or 4 mA DC. Provided that negative readings are allowed (
this is set in the system menu as well ).
If the value is set to zero of FS (full scale), the analog outputs work
in standard form and this is the default.
Summary of Contents for BDS 3000
Page 14: ......
Page 32: ...Operational Theory BDS 3000 Teledyne Analytical Instruments 18...
Page 46: ...Installation BDS 3000 Teledyne Analytical Instruments 32...
Page 76: ...Operation BDS 3000 Teledyne Analytical Instruments 62 Figure 4 2 Analyzer Power up Sequence...
Page 77: ...Oxygen Analyzer Operation Teledyne Analytical Instruments 63 Figure 4 3 Analyzer Span Sequence...
Page 78: ...Operation BDS 3000 Teledyne Analytical Instruments 64 Figure 4 4 Analyzer Zero Sequence...
Page 86: ...Maintenance BDS 3000 Teledyne Analytical Instruments 72...
Page 92: ...Appendix BDS 3000 Teledyne Analytical Instruments 78...
Page 96: ...Appendix MSDS Teledyne Analytical Instruments 82...