Installation
BDS 3000
Teledyne Analytical Instruments
26
Span:
Floating input. 5 to 24 V input across the + and – pins
puts the analyzer into the Span mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows Span mode to terminate when done. A
synchronous signal must open and close external span
valve appropriately. See Figure 3-5 Remote Probe
Connector. (The –C option internal valves operate
automatically.)
Cal Contact:
This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
Table 3-3: Remote Calibration Connections
Pin
Function
9
+ Remote Zero
11
– Remote Zero
10
+ Remote Span
12
– Remote Span
40
Cal Contact
41
Cal Contact
Remote Calibration Protocol:
To properly time the Digital Remote Cal
Inputs to the Model BDS 3000 Analyzer, the customer's controller must
monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote
Cal Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that
request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or
span is complete, the calibration will repeat and the Cal Relay Contact
(CRC) will close again.
Summary of Contents for BDS 3000
Page 14: ......
Page 32: ...Operational Theory BDS 3000 Teledyne Analytical Instruments 18...
Page 46: ...Installation BDS 3000 Teledyne Analytical Instruments 32...
Page 76: ...Operation BDS 3000 Teledyne Analytical Instruments 62 Figure 4 2 Analyzer Power up Sequence...
Page 77: ...Oxygen Analyzer Operation Teledyne Analytical Instruments 63 Figure 4 3 Analyzer Span Sequence...
Page 78: ...Operation BDS 3000 Teledyne Analytical Instruments 64 Figure 4 4 Analyzer Zero Sequence...
Page 86: ...Maintenance BDS 3000 Teledyne Analytical Instruments 72...
Page 92: ...Appendix BDS 3000 Teledyne Analytical Instruments 78...
Page 96: ...Appendix MSDS Teledyne Analytical Instruments 82...