![Teledyne BDS 3000 Operating Instructions Manual Download Page 37](http://html1.mh-extra.com/html/teledyne/bds-3000/bds-3000_operating-instructions-manual_1080476037.webp)
Oxygen Analyzer
Installation
Teledyne Analytical Instruments
23
Fuse Installation:
The fuse block, at the right of the power cord
receptacle, accepts US or European size fuses. A jumper replaces the
fuse in whichever fuse receptacle is not used. Fuses are not installed at
the factory. Be sure to install the proper fuse as part of installation. (See
Fuse Replacement
in chapter 5,
maintenance
.)
3.3.2.2 50-P
IN
E
QUIPMENT
I
NTERFACE
C
ONNECTOR
Figure 3-4 shows the pin layout of the Equipment Interface
connector. The arrangement is shown as seen when the viewer faces the
rear panel of the analyzer. The pin numbers for each input/output
function are given where each function is described in the paragraphs
below.
Figure 3-4: Equipment Interface Connector Pin Arrangement
Analog Outputs:
There are four DC output signal pins—two pins per
output. For polarity, see Table 3-1. The outputs are:
0–1 VDC % of Range:
Voltage rises linearly with increasing
oxygen, from 0 V at 0 ppm to 1 V at
full scale ppm. (Full scale = 100% of
programmable range.)
0–1 VDC Range ID:
0.25 V = Low Range, 0.5 V = Medium
Range, 0.75 V = High Range, 1 V =
100ppm.
4–20 mA DC % Range:
Current increases linearly with
increasing oxygen, from 4 mA at 0
ppm to 20 mA at full scale ppm. (Full
scale = 100% of programmable range.)
4–20 mA dc Range ID:
8 mA = Low Range, 12 mA = Medium
Range, 16 mA = High Range, 20 mA
= 100ppm.
Summary of Contents for BDS 3000
Page 14: ......
Page 32: ...Operational Theory BDS 3000 Teledyne Analytical Instruments 18...
Page 46: ...Installation BDS 3000 Teledyne Analytical Instruments 32...
Page 76: ...Operation BDS 3000 Teledyne Analytical Instruments 62 Figure 4 2 Analyzer Power up Sequence...
Page 77: ...Oxygen Analyzer Operation Teledyne Analytical Instruments 63 Figure 4 3 Analyzer Span Sequence...
Page 78: ...Operation BDS 3000 Teledyne Analytical Instruments 64 Figure 4 4 Analyzer Zero Sequence...
Page 86: ...Maintenance BDS 3000 Teledyne Analytical Instruments 72...
Page 92: ...Appendix BDS 3000 Teledyne Analytical Instruments 78...
Page 96: ...Appendix MSDS Teledyne Analytical Instruments 82...