Operation
BDS 3000
Teledyne Analytical Instruments
52
sufficiently close to zero before pressing
ENTER
to finish
zeroing .
Generally, you have a good zero when sl ope is less than 0.05 ppm/s
for about 30 seconds. When slope is close enough to zero, press
ENT ER.
In
a f ew seconds, the screen will update.
Once zero settling is completed, the information is stored in the
microprocessor, and the instrument automatically returns to the
Analyze
mode.
4.4.2
Span Cal
The
SPAN
button on the front panel is used to span calibrate the
analyzer. Span calibration can be performed using the automatic mode,
where an internal algorithm compares consecutive readings from the
sensor to determine when the output matches the span gas concentration.
Span calibration can also be performed in manual mode, where the
operator determines when the span concentration reading is acceptable
and manually exits the function.
4.4.2.1 A
UTO
M
ODE
S
PANNING
Press
SPAN
to enter the span function. The screen that appears
allows you to select whether the span calibration is to be performed
automatically or manually. Use the
__
arrow keys to toggle between
AUTO
and
MAN
span settling. Stop when
AUTO
appears, blinking, on
the display.
Span: Settling: AUTO
<ENT>
For Next
Press
ENTER
to move to the next screen.
Calib. Holding time
Cal hold: 5 min
This menu allows the operator to set the time the analyzer should
be held in the
AUTO
span mode, after the readings of the analyzer settle.
Five minutes is the default, but it could be adjusted anywhere from 1 to
60 minutes by using the UP or DOWN keys.
Press
ENTER
to move to the next screen.
Span Val: 008.00 ppm
<ENT>Span <UP>Mod #
Summary of Contents for BDS 3000
Page 14: ......
Page 32: ...Operational Theory BDS 3000 Teledyne Analytical Instruments 18...
Page 46: ...Installation BDS 3000 Teledyne Analytical Instruments 32...
Page 76: ...Operation BDS 3000 Teledyne Analytical Instruments 62 Figure 4 2 Analyzer Power up Sequence...
Page 77: ...Oxygen Analyzer Operation Teledyne Analytical Instruments 63 Figure 4 3 Analyzer Span Sequence...
Page 78: ...Operation BDS 3000 Teledyne Analytical Instruments 64 Figure 4 4 Analyzer Zero Sequence...
Page 86: ...Maintenance BDS 3000 Teledyne Analytical Instruments 72...
Page 92: ...Appendix BDS 3000 Teledyne Analytical Instruments 78...
Page 96: ...Appendix MSDS Teledyne Analytical Instruments 82...