Oxygen Analyzer
Maintenance
Teledyne Analytical Instruments
71
situation, uninstall the sensor and take off the cap carefully, then
apply a vacuum degas process as shown in the Figure 5-5.
Degassing in a 28 inch mercury vacuum for 5 minutes is
sufficient to remove the gas bubbles. Reinstall the sensor into the
analyzer.
Note: A low cost vacuum degas kit (TAI P/N B72098) is available
from Teledyne Analytical Instruments.
Figure 5-5: Vacuum Degassing for the BDS Oxygen Sensor