Oxygen Analyzer
Operation
Teledyne Analytical Instruments
59
4.7
The
Analyze
Function
Normally, all of the functions automatically switch back to the
Analyze
function when they have completed their assigned operations.
Pressing the
ESCAPE
button in many cases also switches the analyzer
back to the
Analyze
function. Alternatively, you can press the
ANALYZE
button at any time to return to analyzing your sample.
4.8
Signal Output
The standard Model BDS 3000 Oxygen Analyzer is equipped with
two 0–1 VDC analog output terminals accessible on the back panel (one
concentration and one range ID), and two isolated 4–20 mA DC current
outputs (one concentration and one range ID).
See
Rear Panel
in Chapter 3,
Installation
, for illustration.
The signal output for concentration is linear over the currently
selected analysis range. For example, if the analyzer is set on a range
that was defined as 0–100 ppm O
2
, then the output would be as shown in
Table 4-2.
Table 4-2: Linear Output for a 0-100 ppm O2 Range
Voltage Signal
Current Signal
Summary of Contents for BDS 3000
Page 14: ......
Page 32: ...Operational Theory BDS 3000 Teledyne Analytical Instruments 18...
Page 46: ...Installation BDS 3000 Teledyne Analytical Instruments 32...
Page 76: ...Operation BDS 3000 Teledyne Analytical Instruments 62 Figure 4 2 Analyzer Power up Sequence...
Page 77: ...Oxygen Analyzer Operation Teledyne Analytical Instruments 63 Figure 4 3 Analyzer Span Sequence...
Page 78: ...Operation BDS 3000 Teledyne Analytical Instruments 64 Figure 4 4 Analyzer Zero Sequence...
Page 86: ...Maintenance BDS 3000 Teledyne Analytical Instruments 72...
Page 92: ...Appendix BDS 3000 Teledyne Analytical Instruments 78...
Page 96: ...Appendix MSDS Teledyne Analytical Instruments 82...