4220 Flow Meter
Section 2 Programming
2-39
If you select pH for the parameter to calibrate, the following
display will appear:
You can perform a two- or three-point calibration for pH with the
YSI sonde. The menus that follow are similar to those in the pre-
ceding section for the Isco pH sensor. Select the calibration that
best suits your stream’s profile. When you complete the pH cali-
bration successfully, the following display will appear:
Returning to the YSI menu, if you select D.O:.
ALTITUDE is just off screen to the right. Programming for YSI
600 D.O. is essentially the same as that described for the Isco
D.O. sensor on the preceding section, with the exception that you
always place the sensor in a cup, rather than wrap a moist cloth
around it as is done for the Isco D. O. sensor.
If you select CONDUCTIVITY for the parameter to calibrate, the
following display will appear:
MS/CM is milli-siemens per centimeter. The
siemen
is the S.I.
(
Système Internationale
) name for the unit of conductance, which
is also the reciprocal of the
ohm
. The siemen was formerly called
the
mho
(
ohm
spelled backwards), and that term may be more
familiar to some. PPT is
parts per thousand
. Select the standard
most suitable for your application.
Then:
Then:
If you select PPT for the conductivity standard:
YSI 600 pH CALIBRATION
• pH 4 & 7 • • pH 7 & 10 • • pH 4, 7, & 10 •
CALIBRATING...
PRESS ENTER TO CONTINUE
YSI 600 DISSOLVED OXYGEN CALIBRATION
• D.O. STANDARD• •ABS BAROMETRIC PRESSURE •>
CONDUCTIVITY CALIBRATION UNITS
• MS/CM • • PPT •
CONDUCTIVITY STANDARD
X.XX
MS/CM
PLACE PROBE IN
X.XX
MS/CM
PRESS ENTER WHEN STABLE:
X.XX
MS/CM
CALIBRATING...
PLEASE WAIT...
CONDUCTIVITY STANDARD
X.XX
PPT
Summary of Contents for 4220
Page 2: ......
Page 4: ......
Page 8: ...4220 Flow Meter Safety ...
Page 14: ...vi ...
Page 19: ...4220 Flow Meter Section 1 Introduction 1 5 Figure 1 2 4220 Controls and Indicators ...
Page 28: ...4220 Flow Meter Section 1 Introduction 1 14 ...
Page 94: ...4220 Flow Meter Section 3 Installation 3 12 ...
Page 156: ...4220 Flow Meter Appendix A Replacement Parts A 2 ...
Page 157: ...4220 Flow Meter Appendix A Replacement Parts A 3 ...
Page 160: ...4220 Flow Meter Appendix A Replacement Parts A 6 ...
Page 164: ...4220 Flow Meter Appendix B Accessories List B 4 ...
Page 172: ...4220 Flow Meter Appendix C Programming Worksheets C 8 ...
Page 182: ...4220 Flow Meter Appendix D General Safety Procedures D 10 ...
Page 194: ......
Page 196: ...Warranty ...
Page 198: ......