nickel-cadmium batteries show fairly rapid rates of self-dis-
charge. A battery that is fully charged and then placed in storage
will lose some capacity each day. In a week, this could easily be
5% or more.
When using lead-acid batteries, you must be careful to avoid
complete discharge, as this may cause cell reversal, which will
ruin the battery. Also, complete discharge in low temperature
ambients may cause the battery to freeze, which can deform the
plates or even crack the case. Always operate these batteries
with a reserve factor!
1.7.1 Current Draw
Calculating current draw for a 4220 Flow Meter is somewhat
more difficult than calculating the battery capacity. You cannot
simply measure the idle current of the unit unless the
printer
and
report generator
are turned off in the program. These
functions require power periodically, but not all the time. If the
figures given in the previous table are not satisfactory for your
application, you can use the following procedure (shown in
Figure 1-4) to measure the actual current draw.
Note
Do not attempt this procedure unless you have the proper
equipment available and know how to make electrical mea-
surements.
To measure current for a varying load requires a more-sophisti-
cated type of multimeter, one that is capable of averaging high
and low readings over a period of time. You should set the meter
on MIN/MAX and let it run with your program for several hours
or more. Other manufacturers’ meters are also acceptable, but
only if they are capable of
averaging current draw
. For a more
representative test, the area-velocity sensor should be attached
and submerged in water. You should run the test for at least
eight hours
, longer if necessary, or until the flow meter has
exercised the entire program. The longer you run the test, the
more accurate the average will be.
Summary of Contents for 4220
Page 2: ......
Page 4: ......
Page 8: ...4220 Flow Meter Safety ...
Page 14: ...vi ...
Page 19: ...4220 Flow Meter Section 1 Introduction 1 5 Figure 1 2 4220 Controls and Indicators ...
Page 28: ...4220 Flow Meter Section 1 Introduction 1 14 ...
Page 94: ...4220 Flow Meter Section 3 Installation 3 12 ...
Page 156: ...4220 Flow Meter Appendix A Replacement Parts A 2 ...
Page 157: ...4220 Flow Meter Appendix A Replacement Parts A 3 ...
Page 160: ...4220 Flow Meter Appendix A Replacement Parts A 6 ...
Page 164: ...4220 Flow Meter Appendix B Accessories List B 4 ...
Page 172: ...4220 Flow Meter Appendix C Programming Worksheets C 8 ...
Page 182: ...4220 Flow Meter Appendix D General Safety Procedures D 10 ...
Page 194: ......
Page 196: ...Warranty ...
Page 198: ......