Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
After changing the variable's value, the CPU will respond with:
V VARIABLE=VALUE WARNLO WARNHI [DATALO-DATAHI] <CRLF>
which should reflect the new value. The values in square brackets are not required for all
variables. If needed, the values are included on the command line, separated by spaces. For
example, to change the instrument ID, the host computer would issue a command like this:
V MACHINE_ID=1234<CRLF>
and the CPU should respond with:
V DDD:HH:MM IIII MACHINE_ID = 1234 (0-9999)<CRLF>
Table 7-9 lists the variable names that are variable through the RS-232 interface and their
corresponding button sequences.
Table 7-9: RS-232 Variable Names
Variable Name
Button Sequence
Legal Values
MACHINE_ID
SETUP-MORE-COMM-ID
0000 - 9999
BAUD_RATE
SETUP-MORE-COMM-BAUD
300, 1200, 2400, 4800, 9600, 19.2
CURR_TIME
SETUP-CLK-TIME
00:00 - 23:59
CURR_DATE
SETUP-CLK-DATE
01/01/00 - 12/31/99
7-12
Summary of Contents for 300M
Page 26: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 1 18 ...
Page 36: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 2 10 ...
Page 56: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 4 10 ...
Page 64: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 5 8 ...
Page 66: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 6 2 ...
Page 86: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 8 8 ...
Page 94: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 9 8 ...
Page 116: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 11 4 ...