5 DIAGNOSTICS
The Teledyne API Carbon Monoxide analyzer contains two levels of diagnostics: test
measurements which can be viewed at all times (except when in setup) by pressing
TEST
, and
lower level diagnostic operations which can only be performed by pressing
SETUP-MORE-
DIAG
.
5.1 Test Measurements
As stated, test measurements can be viewed at any time except when in setup. To view a
different test measurement, simply press the
TEST
button. Table 2-3 lists the test measurements
that are available. Viewing these test measurements does not interfere with the operation of the
Model 300M or the Carbon Monoxide reading in any way, so they may be viewed freely.
NOTE
If the value of any test function is displayed as “XXXX”, this
indicates that the reading is off scale or otherwise non-valid.
Additionally, the values of most TEST functions can be output as an analog voltage at the
instrument's rear panel (see Figure 1-3). The TEST function to be output is selected by pressing
SETUP-MORE-DIAG
. Press
NEXT
until TEST CHANNEL OUTPUT appears. Press
ENTR
.
Select test channel function and press
ENTR
. Table 5-1 lists the Test functions available for
analog output.
In addition to outputting a value to the analog output channel, these tests activate a new test
measurement which displays the analog voltage reading on the front panel as:
TEST=XXXXX.X MV
When you exit the diagnostics, this test measurement is removed.
5-1
Summary of Contents for 300M
Page 26: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 1 18 ...
Page 36: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 2 10 ...
Page 56: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 4 10 ...
Page 64: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 5 8 ...
Page 66: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 6 2 ...
Page 86: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 8 8 ...
Page 94: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 9 8 ...
Page 116: ...Teledyne API Model 300M CO Analyzer Instruction Manual 04033 Rev A INTENTIONALLY BLANK 11 4 ...